GENERAL OCEANICS |
Technical Specifications for UCM-60 and UCM-60DL
UCM-60
Current velocity sensors
Range: Programmable:± 3 m/s or: ± 6m/s
Resolution: 1mm/s, 2mm/s at 6m/s range.
Accuracy: Normal operation ± 5mm/s or 1% whichever is greater.
Worst case: 3% of range ± 5 mm/s heavy tilted
Sound velocity sensor
Range: 1380-1580 m/s,
Resolution: 2m/s
Accuracy: 5m/s.
Temperature sensor
Range: -5 C to +45 C
Resolution 0.01 C
Accuracy: 0.1 C
Response: <1s
Option:
Resolution: 0.01 C
Accuracy: 0.02 C
Response: 60 msec at flow rate 1 m/s
Pressure Sensor
Range: from 10 to 350 Bar
Resolution: 0.04% FS
Accuracy: 0.2% FS
Drift: 0.7% TEB (0-50 C)
Response: 0.05 sec.
Conductivity sensor
Range: 2-74 mmho/cm
Resolution: 0.01 mmho/cm
Accuracy: 0.06 mmho/cm
Response: 0.05 sec.
Fluxgate compass
Range: 0-360
Resolution: 1
Accuracy: ±1
Response: 0.05 sec.
Tilt sensor
Range: 0-30
Resolution: 1
Accuracy: 10% of reading
Response: 0.05 sec.
Depth capacity
Depth: 2000m, Option: 6000m
Communication
RS232-C, ASCII, 8 bit, no parity.
Baud rates: 300, 600, 1200, 2400, 4800, and 9600, High speed data upload: 56700 baud.
Housing/material
Cylinder: Stainless steel AISI-316L, Lower part: Stainless steel AISI-316L.
Dimensions (ex. connector)
Cylinder diameter: 102mm
Overall length: ~ 960mm w/sensor protection
Sampling frequency
Data output rate: max. 2 Hz
UCM-60DL
Data logger capacity
Standard: 4 Mbyte
Optional: up to 16 Mbyte
Battery
Standard: Alkaline 15 Ah, Optional Lithium 30 Ah
Power consumption
0.2 - 0.7 W, Depending on programming